UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-04724
- Event Type
- Malfunction
- Date Received
- August 25, 2011
- Date of Event
- July 25, 2011
- Report Date
- July 25, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER INITIALLY REPORTED CALIBRATION FAILURE ON MC (MODULAR CHEMISTRY) ALBUMIN. IT WAS AFTER THE CUSTOMER PERFORMED MAINTENANCE ON THE INSTRUMENT. THE CUSTOMER RAISED THE MODULE TO TROUBLESHOOT AND FOUND IT WAS LEAKING UNDERNEATH. ON (B)(4) 2011, BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT. THE FSE ISOLATED PROBLEM TO THE MC SIDE, AND FOUND A BAD T-VALVE ON MC SYRINGE. THE FSE REPLACED THE T-VALVE. CALIBRATION WAS PERFORMED WITH NO PROBLEMS NOTED. THE FSE VERIFIED REPAIR PER ESTABLISHED PROCEDURES. AS OF (B)(4) 2011, THE CUSTOMER HAS NOT CONTACTED BEC WITH REQUESTS FOR FURTHER ASSISTANCE FOR THIS ISSUE. (B)(4).
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THERE WAS A FLUID LEAK ON UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM. MSDS WAS NOT REVIEWED, BUT THE FACILITY HAS AN EXPOSURE CONTROL PLAN. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION AND THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |