FDA Adverse Event
Malfunction
Summary report: N
FLEXCATH STEERABLE SHEATH
MDR report key: 2222329
·
Received August 24, 2011
Report
- Report Number
- 3002648230-2011-00140
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- July 25, 2011
- Report Date
- July 25, 2011
- Manufacturer
- MEDTRONIC CRYOCATH LP
- Product Code
- DRA
- PMA / PMN Number
- K081049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN (B)(6) 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.
Description of Event or Problem · 1
THE PHYSICIAN SAW A CONSTANT FLOW OF AIR WHEN ASPIRATING FROM THE SIDE PORT OF THE FLEXCATH STEERABLE SHEATH. THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLEXCATH STEERABLE SHEATH | CATHETER, STEERABLE | DRA | MEDTRONIC CRYOCATH LP | 3FC12 | 23512 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |