FLEXCATH STEERABLE SHEATH

MDR report key: 2222329 · Received August 24, 2011

Report

Report Number
3002648230-2011-00140
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
July 25, 2011
Report Date
July 25, 2011
Manufacturer
MEDTRONIC CRYOCATH LP
Product Code
DRA
PMA / PMN Number
K081049
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED ISSUE WAS CONFIRMED THROUGH TESTING. DISSECTION SHOWED A LEAKING HEMOSTATIC VALVE. A FIELD ACTION WAS INITIATED IN (B)(6) 2011 TO COMMUNICATE TO PHYSICIANS AND REGULATORY AUTHORITIES THIS POTENTIAL LEAK IN HEMOSTATIC VALVE OF MEDTRONIC CRYOCATH FLEXCATH 12 STEERABLE SHEATHS.

Description of Event or Problem · 1

THE PHYSICIAN SAW A CONSTANT FLOW OF AIR WHEN ASPIRATING FROM THE SIDE PORT OF THE FLEXCATH STEERABLE SHEATH. THERE WAS NO PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLEXCATH STEERABLE SHEATH CATHETER, STEERABLE DRA MEDTRONIC CRYOCATH LP 3FC12 23512

Patients

Seq Age Sex Outcome Treatment
1