- Report Number
- 6000001-2011-20607
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- July 31, 2011
- Report Date
- August 18, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).THE ASSIGNABLE CAUSE WAS DEPLETED MAIN BATTERIES AS A RESULT OF USE ERROR. THIS ISSUE HAS BEEN ESCALATED TO CAPA.
(B)(4). THE REPORTED CONDITION OF A DAMAGED BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION; AND WAS DETERMINED TO HAVE BEEN CAUSED BY DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |