HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-11204
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- August 3, 2011
- Report Date
- August 3, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE REPORTED SYSTEM ERROR 2240 WAS CONFIRMED. THE ROOT CAUSE WAS INCOMPLETE PRIME. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). PER THE CUSTOMER THE LOT NUMBER WAS UNKNOWN; THEREFORE, NO EVALUATION OR BATCH REVIEW WILL BE PERFORMED. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT HAVING A SYSTEM ERROR 2240 ALARM WITH THE HOMECHOICE (HC) MACHINE DURING INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE AND ADVISED THE HOME PATIENT (HP) TO RECYCLE POWER TO THE HC AND START OVER WITH NEW SUPPLIES. THERE WAS NO REASON FOUND FOR THE ALARM. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. THE HP RETURNED THE CALL MADE BY PRODUCT SURVEILLANCE REGARDING THE REPORTED EVENT. THE HP STATED THE CAUSE OF THE ALARM WAS THAT SHE WAS USING TWO PATIENT LINE EXTENSIONS AND WAS NOT ABLE TO COMPLETELY PRIME THE SETUP. THE HP TRIED TO RE-PRIME THE LINES BUT WAS UNSUCCESSFUL. THE HP ADVISED THAT SHE WAS ABLE TO RESUME THERAPY SUCCESSFULLY WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | HOMECHOICE |