FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2222257 · Received August 24, 2011

Report

Report Number
1423500-2011-11204
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 3, 2011
Report Date
August 3, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED SYSTEM ERROR 2240 WAS CONFIRMED. THE ROOT CAUSE WAS INCOMPLETE PRIME. A LABELING REVIEW OF THE HOMECHOICE APD SYSTEMS PATIENT AT-HOME GUIDE ISSUED WAS FOUND TO BE ADEQUATE FOR THE USE ERROR(S) IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE LOT NUMBER WAS UNKNOWN; THEREFORE, NO EVALUATION OR BATCH REVIEW WILL BE PERFORMED. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER TO REPORT HAVING A SYSTEM ERROR 2240 ALARM WITH THE HOMECHOICE (HC) MACHINE DURING INITIAL DRAIN. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE MESSAGE AND ADVISED THE HOME PATIENT (HP) TO RECYCLE POWER TO THE HC AND START OVER WITH NEW SUPPLIES. THERE WAS NO REASON FOUND FOR THE ALARM. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. THE HP RETURNED THE CALL MADE BY PRODUCT SURVEILLANCE REGARDING THE REPORTED EVENT. THE HP STATED THE CAUSE OF THE ALARM WAS THAT SHE WAS USING TWO PATIENT LINE EXTENSIONS AND WAS NOT ABLE TO COMPLETELY PRIME THE SETUP. THE HP TRIED TO RE-PRIME THE LINES BUT WAS UNSUCCESSFUL. THE HP ADVISED THAT SHE WAS ABLE TO RESUME THERAPY SUCCESSFULLY WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE