FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2222251 · Received August 24, 2011

Report

Report Number
2134265-2011-03200
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
July 8, 2011
Report Date
July 8, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR. THE NC QUANTUM APEX MONORAIL (MR) BALLOON CATHETER WAS RECEIVED WITH NO ORIGINAL PACKAGING. THERE WAS BLOOD AND CONTRAST IN THE HUB AND WIRE LUMEN. INSPECTION OF THE BALLOON REVEALED A PINHOLE IN THE BALLOON WALL ON THE PROXIMAL END OF THE BALLOON. THE SHAFT BETWEEN THE PROXIMAL AND DISTAL MARKERBANDS HAD MULTIPLE KINKS. MAGNIFIED INSPECTION OF THE BALLOON MATERIAL PRESENTED NO INDICATION OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES THAT COULD HAVE CONTRIBUTED TO THE PINHOLE. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY ARTERY INTERVENTIONAL TREATMENT PROCEDURE A BALLOON BURST OCCURRED. THE 90% STENOSED TARGET LESION WAS IN THE MODERATELY TORTUOUS MID LEFT ANTERIOR DESCENDING CORONARY ARTERY. THE PHYSICIAN TREATED THE LESION USING AN NC QUANTUM APEX MR 8MM X 3.50MM BALLOON CATHETER. UPON 4TH INFLATION, THE BALLOON RUPTURED AT UNKNOWN ATM. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO REPORTED COMPLICATION AND THE PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912408350 14053425

Patients

Seq Age Sex Outcome Treatment
1