NC QUANTUM APEX PTCA DILATATION CATHETER
Report
- Report Number
- 2134265-2011-03200
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- July 8, 2011
- Report Date
- July 8, 2011
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- LOX
- PMA / PMN Number
- P860019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATED BY MFR. THE NC QUANTUM APEX MONORAIL (MR) BALLOON CATHETER WAS RECEIVED WITH NO ORIGINAL PACKAGING. THERE WAS BLOOD AND CONTRAST IN THE HUB AND WIRE LUMEN. INSPECTION OF THE BALLOON REVEALED A PINHOLE IN THE BALLOON WALL ON THE PROXIMAL END OF THE BALLOON. THE SHAFT BETWEEN THE PROXIMAL AND DISTAL MARKERBANDS HAD MULTIPLE KINKS. MAGNIFIED INSPECTION OF THE BALLOON MATERIAL PRESENTED NO INDICATION OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES THAT COULD HAVE CONTRIBUTED TO THE PINHOLE. INSPECTION OF THE REMAINDER OF THE DEVICE PRESENTED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS (B)(4).
IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY ARTERY INTERVENTIONAL TREATMENT PROCEDURE A BALLOON BURST OCCURRED. THE 90% STENOSED TARGET LESION WAS IN THE MODERATELY TORTUOUS MID LEFT ANTERIOR DESCENDING CORONARY ARTERY. THE PHYSICIAN TREATED THE LESION USING AN NC QUANTUM APEX MR 8MM X 3.50MM BALLOON CATHETER. UPON 4TH INFLATION, THE BALLOON RUPTURED AT UNKNOWN ATM. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO REPORTED COMPLICATION AND THE PATIENT CONDITION IS GOOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NC QUANTUM APEX PTCA DILATATION CATHETER | CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS | LOX | BOSTON SCIENTIFIC - MAPLE GROVE | H7493912408350 | 14053425 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |