FDA Adverse Event Malfunction Summary report: N

NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM

MDR report key: 2222194 · Received August 24, 2011

Report

Report Number
6000034-2011-00630
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 8, 2011
Report Date
December 8, 2011
Manufacturer
COCHLEAR LTD.
Product Code
MCM
PMA / PMN Number
970051
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
AUDIOLOGIST

Narratives

Additional Manufacturer Narrative · 1

CORRECTION: THE CORRECT DATE OF EXPLANT IS (B)(6) 2011; NOT (B)(6) 2011 AS PREVIOUSLY REPORTED. (B)(4).

Additional Manufacturer Narrative · 1

THIS REPORT IS FILED (B)(4), 2011.

Additional Manufacturer Narrative · 1

(B)(4): IMPLANTED DEVICE REMAINS.

Additional Manufacturer Narrative · 1

PER PATIENT'S SURGEON, THE DEVICE WAS EXPLANTED ON (B)(6), 2011; THE PATIENT WAS REIMPLANTED WITH A NEW DEVICE DURING THE SAME SURGERY. THIS REPORT IS FILED (B)(4), 2011.

Description of Event or Problem · 1

PER THE CLINIC, THE PATIENT EXPERIENCED A SUDDEN LOSS OF CONNECTION TO THE INTERNAL DEVICE. REPROGRAMMING ATTEMPTS WERE MADE; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. THE IMPLANTED DEVICE REMAINS. EXPLANT AND REIMPLANTATION IS PLANNED BUT HAS NOT TAKEN PLACE AT THE TIME OF THIS REPORT (B)(6), 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM MCM, PRODUCT CODE: MCM MCM COCHLEAR LTD. CI512 N/A

Patients

Seq Age Sex Outcome Treatment
1 61 YR Required Intervention