INRATIO
Report
- Report Number
- 2027969-2011-01837
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Date of Event
- July 18, 2011
- Report Date
- August 17, 2011
- Manufacturer
- ALERE SAN DIEGO, INC
- Product Code
- GJS
- PMA / PMN Number
- 021923
- Removal / Correction Number
- NI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INRATIO PRECISION DATA PROVIDED BY END-USER: 1ST INR: 1.1; 2ND INR: 3.3; MEAN: 2.20; SD: 1.56; %CV: 70.71. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT REPEATED INRATIO TEST RESULT COMPARISON DID NOT MEET PRECISION CRITERIA. CUSTOMER'S RESULTS ARE CONSIDERED DISCREPANT BEYOND THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. RETAINED STRIP TESTING PERFORMED ON 07/0/2011 REVEALED THAT RESULT COMPARISONS MET PRECISION CRITERIA. INVESTIGATION RESULTS FOR RETAINED STRIP LOT #243104: DONOR 74: 1ST INR: 2.1; 2ND INR: 1.9; 3RD INR: 2.0; MEAN: 2.00; %CV: 5.00. DONOR 75 - 1ST INR: 3.6; 2ND INR: 3.7; 3RD INR: 3.7; MEAN: 3.67; %CV: 1.57. %CV FOR BOTH DONORS ARE LESS THAN 16% AND MEET THE CRITERIA FOR PRECISION. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. (B)(4). ACTION THRESHOLD (B)(4) WAS REACHED. STRIP LOT IN COMPLAINT HAS STRIP LOT CODE Z45BU WITH BRICK LOT #246544 WHICH WAS SPLIT INTO TWO DIFFERENT LOTS (243104 AND 251117). (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE TOTAL ACTION THRESHOLD OF (B)(4), NO FURTHER ACTION IS REQUIRED. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CUSTOMER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: DATE: (B)(6) 2011; INRATIO: 1.1, 3.3. PT'S THERAPEUTIC RANGE IS 2.0-3.0 INR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC | 100071 | 243104 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |