FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2222044 · Received August 17, 2011

Report

Report Number
3004209178-2011-82587
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
July 24, 2011
Report Date
July 24, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF CORRODED INTERFACE BOARD AND MOTHERBOARD. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE FROZEN DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP HAD AN ERROR ALARM. THE CUSTOMER COULD NOT CLEAR THE ALARM, AND THE BUTTONS WERE UNRESPONSIVE. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-515PNAS

Patients

Seq Age Sex Outcome Treatment
1 41 YR