FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX

MDR report key: 2222019 · Received August 17, 2011

Report

Report Number
1218950-2011-02413
Event Type
Malfunction
Date Received
August 17, 2011
Report Date
July 21, 2011
Manufacturer
PHILIPS HEALTHCARE - ANDOVER
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER REPORTED THAT THE DEVICE WOULD LOCK UP AT THE CHARGE (ENERGY) STEP. THERE WAS NO REPORT OF PT INVOLVEMENT. THE PHILIPS CUSTOMER REPAIR CENTER EVALUATED THE DEVICE AND CONFIRMED THE MALFUNCTION. THE DEVICE WOULD LOCK UP DURING THE BEGINNING OF THE CHARGE TEST SEGMENT OF THE USER INITIATED OPERATIONAL CHECK. THE MALFUNCTION WAS RESOLVED BY RELOADING THE SOFTWARE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE DEVICE WOULD LOCK UP AT THE CHARGE (ENERGY) STEP. THERE WAS NO REPORT OF PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTSTART MRX MKJ PHILIPS HEALTHCARE - ANDOVER M3535A