FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX
MDR report key: 2222019
·
Received August 17, 2011
Report
- Report Number
- 1218950-2011-02413
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Report Date
- July 21, 2011
- Manufacturer
- PHILIPS HEALTHCARE - ANDOVER
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED THAT THE DEVICE WOULD LOCK UP AT THE CHARGE (ENERGY) STEP. THERE WAS NO REPORT OF PT INVOLVEMENT. THE PHILIPS CUSTOMER REPAIR CENTER EVALUATED THE DEVICE AND CONFIRMED THE MALFUNCTION. THE DEVICE WOULD LOCK UP DURING THE BEGINNING OF THE CHARGE TEST SEGMENT OF THE USER INITIATED OPERATIONAL CHECK. THE MALFUNCTION WAS RESOLVED BY RELOADING THE SOFTWARE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE DEVICE WOULD LOCK UP AT THE CHARGE (ENERGY) STEP. THERE WAS NO REPORT OF PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS HEALTHCARE - ANDOVER | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |