FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 2222010 · Received August 17, 2011

Report

Report Number
1220908-2011-02233
Event Type
Malfunction
Date Received
August 17, 2011
Report Date
July 27, 2011
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K990762
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE'S BATTERY MELTED DOWN AND WOULD NOT ALLOW THE DEVICE TO POWER UP. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA