FDA Adverse Event Injury Summary report: N

PROXIMATE ILS CURVED CIR STAPL

MDR report key: 2221894 · Received August 24, 2011

Report

Report Number
3005075853-2011-03486
Event Type
Injury
Date Received
August 24, 2011
Date of Event
July 20, 2011
Report Date
August 2, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STOMACH RESECTION PROCEDURE, JEJUNUM-OESOPHAGUS ANASTOMOSIS, AFTER THE SURGEON FIRED THE DEVICE, THE TISSUE COULD NOT BE CUT COMPLETELY. THEN THE SURGEON USED KNIFE TO CUT. BUT THE ANASTOMOSIS HAD TRAUMA. THE SURGEON OPENED THE CHEST TO ANASTOMOSE. THE SURGERY WAS DELAYED TWO HOURS. NOW THE PATIENT IS FINE. ADDITIONAL INFORMATION BEING REQUESTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS REQUESTED ON (B)(4) 2011 AND TO DATE, NO MORE INFORMATION HAS BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE ILS CURVED CIR STAPL STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4RU2V

Patients

Seq Age Sex Outcome Treatment
1 64 YR