FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2221821 · Received August 24, 2011

Report

Report Number
2531779-2011-06178
Event Type
Malfunction
Date Received
August 24, 2011
Report Date
July 26, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE UP AND DOWN ARROW BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES AND REQUIRED INCREASED FORCE TO ENGAGE. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

A FAMILY MEMBER REPORTED THAT THE UP ARROW BUTTONS WERE REALLY HARD TO GET TO WORK; SHE WAS UNSURE ON HOW LONG THE PUMP HAD THIS ISSUE. THE PATIENT REPORTEDLY WORE THE PUMP IN A PUMP CASE IN A POCKET AND CLEANED THE PUMP WITH A DAMP PAPER TOWEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 13 YR