FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2221771 · Received August 17, 2011

Report

Report Number
1824206-2011-04344
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON INSPECTION, THE TECH FOUND THAT THE BALL SCREW ASSEMBLE WAS DRY AND VERY DIRTY AND THE MOTOR COUPLER WAS CRACKED. HE CLEANED AND LUBRICATED THE BALL SCREW ASSEMBLY AND REPLACED THE MOTOR COUPLER, BUT THE BED CONTINUED TO DRIFT. HE REPLACED THE BALL SCREW ASSEMBLY TO REPAIR THE BED.

Description of Event or Problem · 1

THE ACCOUNT STATED THAT THE BED WAS NOISY AND WHEN RAISED, WOULDN'T STAY UP ALL THE TIME, IT WOULD DRIFT DOWN SLOWLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1