FDA Adverse Event Malfunction Summary report: N

KEEPSAFE DELUXE NO POWER SWITCH

MDR report key: 2221738 · Received August 17, 2011

Report

Report Number
2020362-2011-00268
Event Type
Malfunction
Date Received
August 17, 2011
Report Date
July 18, 2011
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVAL: RESULTS: EVAL OF THE RETURNED PRODUCT FOUND THAT THERE IS CORROSION IN THE BATTERY COMPARTMENT. THE AQUA-ALERT WATER INDICATOR LABEL SHOWS MOISTURE EXPOSURE. UNIT POWER UP AND PASSES ALL FUNCTIONAL TESTS. NOTE: THE IFU HAS A WARNING STATEMENT: THE POSEY KEEPSAFE DELUXE IS AN ELECTRONIC DEVICE. IT MAY FAIL TO WORK IF SUBJECTED TO SEVERE SHOCK, SUCH AS BEING DROPPED, OR IMMERSED IN LIQUID. TO REDUCE THE RISK OF SERIOUS INJURY OR DEATH, TEST THE ALARM AND SENSOR FOR PROPER OPERATION PRIOR TO PUTTING IN SVC WITH A PT, AND EACH TIME BEFORE LEAVING THE PT UNATTENDED. USE OF MIXED BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. REMOVE ANY ALARM FROM USE AND SEND TO THE APPROPRIATE FACILITY AUTHORITY IF BATTERIES ARE DAMAGED OR CORRODED OR THE BATTERY COMPARTMENT HAS SIGNS OF PREVIOUS BATTERY CORROSION SUCH AS WHITE POWDER RESIDUE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY HAVE BEEN USING THE ALARM WITH THE ORIGINAL BATTERIES FOR ONLY A MONTH. THE ALARM NOW HAS BATTERY LEAKAGE AND CORROSION. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KEEPSAFE DELUXE NO POWER SWITCH KMI J. T. POSEY CO. 8374NP NA

Patients

Seq Age Sex Outcome Treatment
1 NI