FDA Adverse Event Malfunction Summary report: N

KEEPSAFE DELUXE NO POWER SWITCH

MDR report key: 2221737 · Received August 17, 2011

Report

Report Number
2020362-2011-00267
Event Type
Malfunction
Date Received
August 17, 2011
Report Date
July 18, 2011
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - CORROSION. EVAL RESULTS: EVAL OF THE RETURNED PRODUCT FOUND CORROSION ON THE BATTERY SPRINGS. THE ALARM POWERS UP BUT DOES NOT RETAIN ITS PREVIOUS SETTINGS WHEN POWERED OFF AND POWERED ON. NOTE: POSEY'S IFU WARNING STATEMENT FOR BATTERY INSTALLATION: TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. ALWAYS INSTALL A COMPLETELY NEW SET OF BATTERIES WHEN THE CHIRPING DUE TO LOW BATTERY POWER SOUND BEGINS. DO NOT REPLACE A SINGLE CELL, BUT ALL THE CELLS IN THE ALARM. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK (4 BATTERIES). USE OF MIXED BATTERIES OR BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. DO NOT USE THE ALARM IF ANY EVIDENCE OF BATTERY ACID IS NOTED IN THE BATTERY COMPARTMENT. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY HAVE BEEN USING THE ALARM WITH THE ORIGINAL BATTERIES FOR ONLY A MONTH. THE ALARM NOW HAS BATTERY LEAKAGE AND CORROSION. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KEEPSAFE DELUXE NO POWER SWITCH KMI J. T. POSEY CO. 8374NP NA

Patients

Seq Age Sex Outcome Treatment
1 NI