COLLEAGUE
Report
- Report Number
- 6000001-2011-20543
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- August 5, 2011
- Report Date
- August 8, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH CHARGE LIGHT FLICKERING WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE FAULTY UIM PCB (USER INTERFACE MODULE PRINTED CIRCUIT BOARD). APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE UIM PCB. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.04.00 WHICH IS CATEGORIZED AS UNREMEDIATED.
THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH THE CHARGING LIGHT IS FLICKERING. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |