FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2221668 · Received August 24, 2011

Report

Report Number
6000001-2011-20543
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 5, 2011
Report Date
August 8, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH CHARGE LIGHT FLICKERING WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THE REPORTED PROBLEM WAS DETERMINED TO BE FAULTY UIM PCB (USER INTERFACE MODULE PRINTED CIRCUIT BOARD). APPROPRIATE ACTION WAS TAKEN TO CORRECT THE REPORTED PROBLEM BY REPLACING THE UIM PCB. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.04.00 WHICH IS CATEGORIZED AS UNREMEDIATED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP IN WHICH THE CHARGING LIGHT IS FLICKERING. THIS CONDITION HAS THE POTENTIAL TO CAUSE AN INTERRUPTION OF DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.