FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2221641 · Received August 24, 2011

Report

Report Number
1423500-2011-11180
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
July 13, 2011
Report Date
August 3, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE PAL EVALUATED THE DEVICE AND NO FAILURE OR MALFUNCTION WAS IDENTIFIED THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) FOUND IN THE DEVICE LOGS. THE CAUSE OF THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) WAS DETERMINED TO BE: INSUFFICIENT DRAIN, FALSE EMPTY DETECT AND USE ERROR, INITIAL DRAIN ALARM SETTING INAPPROPRIATELY PROGRAMMED TOO LOW. DEVICE MET SPECIFICATIONS RELATIVE TO IIPV IN THE LOGS. THERE WERE NO PROBLEMS FOUND DURING AN INTERNAL INSPECTION OF THE DEVICE THAT WOULD CONTRIBUTE TO THE IIPV. A REVIEW OF THE DEVICE SERVICE HISTORY FOR THIS PARTICULAR HOMECHOICE REVEALED THE PREVIOUS SWAP WAS UNRELATED. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE EVALUATION IS IN PROCESS. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

DURING INITIAL ASSESSMENT OF A RETURNED HOMECHOICE MACHINE, A BAXTER TECHNICIAN FOUND AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT THAT WAS IDENTIFIED IN THE PATIENT THERAPY LOG. THE DRAIN VOLUME WAS 4950 ML DURING CYCLE 1. THE FILL VOLUME WAS 2500 ML. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED. THIS EVENT MEETS OVERFILL CRITERIA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1