FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 2221640 · Received August 24, 2011

Report

Report Number
1423500-2011-11179
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 3, 2011
Report Date
August 3, 2011
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR, BECAUSE, THE PRODUCT IS UNKNOWN AT THIS TIME. THE SAMPLE WAS NOT RETURNED TO BAXTER, THEREFORE, NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR THE REPORTED ISSUES OF CHECK PATIENT LINE ALARM AND "LINE FULL OF AIR" WAS NOT CONFIRMED AND A CAUSE IS UNDETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA.

Description of Event or Problem · 1

DURING TROUBLESHOOTING OF A CHECK PATIENT LINE ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING INITIAL DRAIN, THE HOME PATIENT (HP) REVEALED THAT THE LINE APPEARED TO BE FULL OF AIR. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO END THERAPY AND START OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE