FDA Adverse Event Injury Summary report: N

PFNA 10 130 DEGREE L240 TAN

MDR report key: 2221606 · Received August 19, 2011

Report

Report Number
8030965-2011-00558
Event Type
Injury
Date Received
August 19, 2011
Report Date
August 1, 2011
Manufacturer
SYNTHES GMBH
Product Code
HSB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
SERVICE PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION COULD NOT BE COMPLETED, NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RETURNED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. NOT PREVIOUSLY REPORTED. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO A DEVICE MARKETED IN THE USA. MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. EXPIRATION DATE NOT PREVIOUSLY REPORTED. THE MEASURABLE DIMENSIONS OF THE PFNA HAVE BEEN CHECKED. THE ARTICLE CORRESPONDS WITH THE DRAWINGS AND AO/ASIF SPECIFICATIONS. THE INSPECTION OF THE RAW MATERIAL TEST CERTIFICATE AND THE MANUFACTURING DOCUMENTS SHOW NO DEVIATION ACCORDING TO MATERIAL ANALYSIS, TENSILE STRENGTH AND STRUCTURAL STABILITY AS WELL AS OF THE PRODUCTION. THE VALUE CONFORMS TO THE AO/ASIF SPECIFICATIONS AND INTERNATIONAL STANDARDS ISO 5832-11. THE INVESTIGATION OF THE COMPLAINED PFNA NAIL SHOWS A BREAK THROUGH THE DISTAL LOCKING HOLE. ON THE BASIS OF THE BREAK AND OF THE HOMOGENEOUS BREAK CHARACTERISTIC WE ASSUME THAT OVER-STRESSING OF THE IMPLANT WAS LEADING TO THE FAILURE. NO PRODUCT FAULT COULD BE DETECTED.

Description of Event or Problem · 1

THIS REPORT IS #1 OF 1 FOR COMPLAINT (B)(4).

Description of Event or Problem · 1

A DEVICE REPORT FROM (B)(6) INDICATES A HOSPITAL IN (B)(6) REPORTED: A PFNA NAIL 130 DEGREE WAS DISCOVERED BROKEN IN THE SHAFT (DISTAL LOCKING BOLT JUNCTION) AND MEDICAL/SURGICAL INTERVENTION WAS NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PFNA 10 130 DEGREE L240 TAN PFNA 10 130 DEGREE HSB SYNTHES GMBH 2678852

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention DISTAL LOCKING BOLT