FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 2221463
·
Received August 17, 2011
Report
- Report Number
- 3008642652-2011-00451
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Date of Event
- March 28, 2011
- Report Date
- August 12, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BROKEN TRUNK CONNECTOR) HAS BEEN CONFIRMED. UPON EVAL, THE TRUNK CABLE CONNECTOR WAS DAMAGED. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE BROKEN CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY PROBLEMS.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE EVENT. DURING SERVICING OF ELECTRODE BELT SN (B)(4) WHICH WAS RETURNED FOR ROUTINE MAINTENANCE, IT WAS DISCOVERED THAT THE BELT HAD A BROKEN TRUNK CABLE CONNECTOR. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY PROBLEM WITH IT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |