FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2221456 · Received August 17, 2011

Report

Report Number
3008642652-2011-00464
Event Type
Malfunction
Date Received
August 17, 2011
Date of Event
June 25, 2011
Report Date
August 16, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CODE 202) HAS BEEN CONFIRMED. THE CAUSE OF THE CODE 202 IS CORRUPTED FLASH PROGRAMMING. THE ROOT CAUSE OF THE CORRUPTED FLASH CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE CORRUPTED FLASH. THE PT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) DISTRIBUTOR RETURNED A MONITOR THAT WAS GENERATING A CODE 202.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA