LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00456
- Event Type
- Malfunction
- Date Received
- August 17, 2011
- Date of Event
- July 1, 2011
- Report Date
- September 15, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER ON) HAS BEEN CONFIRMED. THE CAUSE OF THE MONITOR NOT STARTING UP WAS A FLASH MEMORY FAILURE (COMPONENTS U102 AND U105) ON THE C/A BOARD. AN INTERMITTENT CONNECTION WAS DISCOVERED AT PIN A25 OF THE FLASH MEMORY. THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURE SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MEMORY. THE PT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A LOUD HUMMING NOISE COMING FROM HIS MONITOR. WHEN CUSTOMER SUPPORT ASKED THE PT TO REMOVE AND RE-INSERT THE BATTERY, THE MONITOR WOULD NOT POWER ON. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |