CARDIOCHEK PLUS ANALYZER
Report
- Report Number
- 1836135-2025-00009
- Event Type
- Malfunction
- Date Received
- June 16, 2025
- Date of Event
- February 24, 2025
- Report Date
- June 16, 2025
- Manufacturer
- PTS DIAGNOSTICS
- Product Code
- CGA
- UDI-DI
- 00382932700016
- PMA / PMN Number
- K140068
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
UPON VISUAL INSPECTION OF THE BATTERY COMPARTMENT, PLASTIC CASE DAMAGE INDICATIVE OF IMPROPER/FORCEFUL REMOVAL OF THE BATTERIES WITH A FOREIGN OBJECT WAS OBSERVED. PTS IS CONFIDENT THAT THIS DAMAGE OCCURRED POST-DISTRIBUTION, AS THERE ARE NO MANUFACTURING STEPS THAT WOULD CAUSE THIS TYPE OF DAMAGE. THE CUSTOMER ALLEGATION OF OVERHEATING WAS NOT OBSERVED ON THE RETURNED ANALYZER. NUMEROUS BATTERY COMPARTMENT IMPROVEMENTS HAVE ALREADY BEEN IMPLEMENTED PRIOR TO THESE ALLEGATIONS FROM THE FIELD. PROCESS IMPROVEMENTS AND PART SPECIFICATIONS HAVE ENHANCED SEATING AND HOLD CAPACITY OF THE BATTERY CONTACTS. ADDITIONALLY, A NEW BATTERY COMPARTMENT DESIGN (APPROVED BY THE FDA PER K193406) ELIMINATES THE POTENTIAL FOR OVERHEATING BY REMOVING THE POINT OF CONTACT THAT AN UNSEATED TERMINAL WOULD REACH.
THE CONTACT REPORTED THAT THEIR ANALYZER GOT WARM UPON BATTERY INSTALLATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1040302 | CARDIOCHEK PLUS ANALYZER | CLINICAL CHEMISTRY ANALYZER | CGA | PTS DIAGNOSTICS | 1.11 | 00382932700016 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |