OLYMPUS SONOSURG SCISSORS 5 MM O.D., HF SERIES
Report
- Report Number
- 8010047-2011-00186
- Event Type
- Malfunction
- Date Received
- August 12, 2011
- Date of Event
- April 27, 2011
- Report Date
- July 14, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- LFL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
OLYMPUS FOLLOWED-UP WITH THE USER FACILITY TO OBTAIN ADD'L INFO REGARDING THIS REPORT. THE USER FACILITY REPORTED THAT THE PT HAD BEEN DISCHARGED FROM THE HOSPITAL ON (B)(6) 2011 AND WAS REPORTEDLY DOING FINE. THE BROKEN DEVICE REPORTEDLY DID NOT HAVE ANY SHARP EDGES AND WAS 0.8MM IN DIAMETER AND 0.7MM IN LENGTH. THE DEVICE REFERENCED IN THIS REPORT WAS RETURNED TO THE ORIGINAL EQUIPMENT MFR (OEM) FOR EVAL. THE EVAL FOUND ONE OF THE PINS ON THE GRASPING PORTION OF THE DEVICE WAS BROKEN AND THE OTHER PIN WAS BENT. THE OEM CONCLUDED THAT THE DISTAL END OF THE SUBJECT DEVICE BROKE DUE TO EXCESSIVE FORCE APPLIED TO THE GRASPING PORTION OF THE DEVICE IN THE ATTEMPT TO OPENING IT FURTHER. THE USER FACILITY REPORTED THAT THE DEVICE HAD BEEN USED EXTENSIVELY. THE MFG HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. THIS REPORT IS BEING SUBMITTED AS AN MDR IN AN ABUNDANCE OF CAUTION.
OLYMPUS WAS INFORMED THAT DURING A RESECTION OF THE THYMUS FOR THYMOMA PROCEDURE, THE DISTAL END OF THE SUBJECT DEVICE WAS NOTED BROKEN ON THE ENDOSCOPIC IMAGE. THE PROCEDURE WAS REPORTEDLY COMPLETED WITH A DIFFERENT, UNIDENTIFIED DEVICE. A CHEST X-RAY AND CT SCAN WERE SAID TO HAVE BEEN PERFORMED, BUT COULD NOT LOCATE THE DEVICE. THE USER FACILITY HAD REPORTEDLY DETERMINED THAT THE BROKEN PORTION OF THE DEVICE HAD FALLEN OFF OUTSIDE OF THE PT'S BODY. THERE WAS NO PT HARM REPORTED. THE PT WAS REPORTEDLY DISCHARGED AND IS DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS SONOSURG SCISSORS 5 MM O.D., HF SERIES | SONOSURG SCISSORS | LFL | OLYMPUS MEDICAL SYSTEMS CORP. | T3105 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |