DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2221348 · Received August 24, 2011

Report

Report Number
1226181-2011-00053
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
July 27, 2011
Report Date
July 27, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
JJE
PMA / PMN Number
K010061
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY DEPRESSED SODIUM RESULT WAS MULTIPLE MAINTENANCE ISSUES INCLUDING USE OF AN EXPIRED ACCESSORY FLUID. THE SIEMENS HEALTHCARE DIAGNOSTICS TECHNICAL SERVICE CENTER REPRESENTATIVE DIRECTED THE CUSTOMER TO APPROPRIATE MAINTENANCE PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY DEPRESSED SODIUM RESULT WAS OBTAINED ON A PATIENT SAMPLE. IT IS UNKNOWN IF THE RESULT WAS REPORTED TO THE PHYSICIAN. AFTER DISCORDANCE WAS RECOGNIZED, THE SAMPLE WAS RE-RUN AND A HIGHER RESULT WITHIN THE NORMAL RANGE WAS OBTAINED AND REPORTED. PATIENT TREATMENT WAS NOT REPORTED AS ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY DEPRESSED SODIUM RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM JJE SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW XPAND HM

Patients

Seq Age Sex Outcome Treatment
1