GYNECARE X-TRACT MORCELLATOR
MDR report key: 2221346
·
Received August 24, 2011
Report
- Report Number
- 2210968-2011-01163
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- June 15, 2011
- Report Date
- August 11, 2011
- Manufacturer
- ETHICON INC
- Product Code
- HET
- PMA / PMN Number
- K993801
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE CPC CONNECTOR WAS FOUND TO BE MISSING THE CLIP, TWO SPRINGS, AND A DOWEL PIN. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, A PIECE BROKE OFF OF THE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.