GYNECARE X-TRACT MORCELLATOR
MDR report key: 2221344
·
Received August 24, 2011
Report
- Report Number
- 2210968-2011-01161
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- June 7, 2011
- Report Date
- August 11, 2011
- Manufacturer
- ETHICON INC
- Product Code
- HET
- PMA / PMN Number
- K993801
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE FLEX COUPLER WAS FOUND TO MOVE FREELY ON THE SHAFT DUE TO A BROKEN SCREW SET. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, THE DEVICE WAS MAKING A RATTLING NOISE. THE CASE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.