FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 2221344 · Received August 24, 2011

Report

Report Number
2210968-2011-01161
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
June 7, 2011
Report Date
August 11, 2011
Manufacturer
ETHICON INC
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE FLEX COUPLER WAS FOUND TO MOVE FREELY ON THE SHAFT DUE TO A BROKEN SCREW SET. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2011. DURING THE PROCEDURE, THE DEVICE WAS MAKING A RATTLING NOISE. THE CASE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON INC NA NA

Patients

Seq Age Sex Outcome Treatment
1