FDA Adverse Event Malfunction Summary report: N

INFUSO.R.

MDR report key: 2221312 · Received August 24, 2011

Report

Report Number
6000001-2011-20474
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 2, 2011
Report Date
August 5, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K883577
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RETURNED FOR EVALUATION HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. ONCE THE EVALUATION IS COMPLETE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF "NEEDS A NEW SWITCHBOARD" INVOLVING AN INFUSO.R. PUMP WAS CONFIRMED AND REPRODUCED DURING DEVICE EVALUATION. THE ASSIGNABLE CAUSE WAS DETERMINED TO BE A DAMAGED SWITCHBOARD DUE TO A DEFECTIVE SWITCH PRINTED CIRCUIT BOARD. THE SWITCHBOARD WAS REPLACED TO FIX THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED AN INFUSO.R. PUMP WHICH WOULD NOT BOLUS DUE TO AN INOPERABLE SWITCH BOARD. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT DELIVERY. THE PROCESS STEP IS UNKNOWN. IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, PATIENT INJURY, OR MEDICAL INTERVENTION ACCORDING TO THE HOSPITAL REPRESENTATIVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSO.R. PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1