PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2011-82584
- Event Type
- Injury
- Date Received
- August 17, 2011
- Date of Event
- August 3, 2011
- Report Date
- August 4, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF OVER 600 MG/DL. THE CUSTOMER STATED THAT THE LAST INFUSION SET CHANGE PRIOR TO HIS ADMISSION WAS (B)(6), 2011. THE CUSTOMER WAS EXPERIENCING UNEXPLAINED HIGH GLUCOSE FOR THE PAST TWO DAYS. THE CUSTOMER HAS TREATED HIS HIGH GLUCOSE LEVEL WITH THE INSULIN PUMP AND MANUAL INJECTIONS. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING, TIME, AND DATE WERE CORRECT. RAN A FIXED PRIME AND HIGH PRESSURE TEST AND THE INSULIN PUMP PASSED THE TESTS. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization | PARADIGM: 9MM CATHETER| QUICK-SET: 23| UNOMEDICAL INFUSION SET: MMT-397 |