FDA Adverse Event Injury Summary report: N

COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER

MDR report key: 2221247 · Received August 17, 2011

Report

Report Number
3002808486-2011-00029
Event Type
Injury
Date Received
August 17, 2011
Report Date
July 15, 2011
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
PMA / PMN Number
K061815
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION FROM AUGUST 08, 2011, IS ONLY BASED ON THE DESCRIPTION ONLY, SINCE IMAGES AND PRODUCT HAVE NOT BEEN AVAILABLE. (B)(4). THE EVAL OF THIS COMPLAINT WAS BASED ON THE DESCRIPTION ONLY SINCE NO IMAGES HAVE BEEN AVAILABLE. WITHOUT CT SCANS, IT IS NOT POSSIBLE TO DETERMINE IF THE FILTER PERFORATES THE IVC OR IF THE FILTER IS JUST IN A TENTING SITUATION. BASED ON A REVIEW OF THE CLINICAL LITERATURE, COOK HAS CONCLUDED THAT VENA CAVA FILTERS DISTORT THE VESSEL LUMEN SO THAT THE ANCHORING POINTS OF THE FILTER APPEARS OUTSIDE THE BLOOD FLOW FLUOROSCOPICALLY VISUALIZED. MOREOVER, THIS DISTORTION OR "TENTING" IS NOT UNIQUE TO THE CELECT FILTER. TENTING IS A GENERIC PROPERTY OF THIS CLASS OF DEVICE. HOWEVER, FILTER PERFORATION OF THE VENA CAVA WALL IS A WELL KNOWN RISK. SEVERAL CASE REPORTS PUBLISHED IN ARTICLES, DESCRIBE FILTER PERFORATION OF THE VENA CAVA WALL. DESPITE PERFORATION, THEY ALL END UP IN SUCCESSFUL FILTER RETRIEVAL. NO OTHER COMPLAINTS RECEIVED ON THIS LOT NUMBER.

Description of Event or Problem · 1

THE PT HAD A COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER PLACED IN (B)(6) 2010. DURING A FOLLOW-UP VISIT IN (B)(6) 2011, AN ABDOMEN PELVIS CT SCAN DEMONSTRATED THE FILTER IS NOTED IN THE IVC AND FOUR OF THE WIRES HAVE PASSED OUTSIDE THE IVC AS PRESENT ON THE PRIOR STUDY WITH A PRONG THROUGH THE AORTA, DUODENUM AND INTO THE L4 VERTEBRAL BODY. THE PT HAS NO DISCOMFORT, AND IS STABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE NA E2281675

Patients

Seq Age Sex Outcome Treatment
1 UNK Life Threatening