FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2221209 · Received August 24, 2011

Report

Report Number
1423500-2011-11170
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED BY THE CUSTOMER. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED, AND THE ROOT CAUSE COULD NOT BE DETERMINED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TECHNICAL SERVICES (BTS) REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 ALARM DURING DWELL 2 ON THE HOMECHOICE MACHINE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HOME PATIENT (HP) TO CYCLE POWER TO GET TO "PRESS GO TO START". THE TSR ADVISED THE HP TO START OVER WITH NEW SUPPLIES. THE HP WAS CONTACTED ON (B)(6) 2011. THE HP STATED THIS HAS ONLY OCCURRED ONCE PRIOR. THE HP STATED THAT HE DID NOT DEVELOP ANY ADVERSE REACTIONS OR NEED ANY MEDICAL INTERVENTION. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 56 YR HOMECHOICE PRO