TAXUS (TM) LIBERTÉ (TM)

MDR report key: 2221190 · Received August 24, 2011

Report

Report Number
2134265-2011-03558
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 7, 2011
Report Date
August 7, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE IDENTIFIED THAT THE TIP OF THE DEVICE WAS SLIGHTLY FLARED. STENT STRUTS FROM THE MOST PROXIMAL ROW WERE BENT OUTWARDS DISTALLY. THE BALLOON SECTION OF THE DEVICE WAS VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH ITS PROFILE THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. SOLIDIFIED BLOOD WAS PRESENT ON THE OUTSIDE OF THE BALLOON. THE HYPOTUBE WAS KINKED ALONG ITS ENTIRE LENGTH. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT AND BALLOON DAMAGE OCCURRED. THE TARGET LESION BEING TREATED WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 4.00X16MM TAXUS LIBERTE STENT DELIVERY SYSTEM WAS PREPPED AND FOUND TO HAVE PROXIMAL STENT DAMAGE AND A KINK IN THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT AND BALLOON DAMAGE OCCURRED. THE TARGET LESION BEING TREATED WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A 4.00X16MM TAXUS LIBERTE STENT DELIVERY SYSTEM WAS PREPPED AND FOUND TO HAVE PROXIMAL STENT DAMAGE AND A KINK IN THE BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS (TM) LIBERTÉ (TM) CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894016450 14299690

Patients

Seq Age Sex Outcome Treatment
1 64 YR