AU2703-03E CLINICAL CHEMISTRY ANALYZER.
Report
- Report Number
- 2050012-2011-04810
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- July 28, 2011
- Report Date
- July 28, 2011
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K003721
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SAMPLES WERE FROM FRESH ADULT PATIENT SAMPLES AND ANALYZED FROM THE PRIMARY TUBES WHICH WERE VACUETTE BD 4 ML TUBES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE SAMPLE PROBE AND THE ISE SAMPLE D-POT ON THE ISE UNIT AS THE SAMPLE ARM, WHICH WAS PREVIOUSLY REPLACED DUE TO MISALIGNMENT, WAS CAUSING GROVES IN THE SAMPLE D-POT. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT AFTER SERVICE WAS PERFORMED FOR A PREVIOUS REPORTABLE EVENT (MDR 2050012-2011-04801) THE (B)(4) CLINICAL CHEMISTRY ANALYZER GENERATED ERRONEOUSLY LOW SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) RESULTS ON TWO (2) PATIENT SAMPLES. THE TWO (2) SAMPLES WERE THEN REPEATED AUTOMATICALLY ON THE SAME INSTRUMENT AND GAVE HIGHER RECOVERIES. THE SAMPLES WERE THEN TRANSFERRED TO ANOTHER INSTRUMENT IN THE LABORATORY AND CONFIRMED THE REPEAT RUN VALUES AND THE RESULTS WERE THEN REPORTED OUT OF THE LABORATORY. NO TREATMENT TO PATIENT WAS GIVEN UNTIL RESULTS WERE RELEASED AFTER RETESTING. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU2703-03E CLINICAL CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | AU2701-03ISE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |