FDA Adverse Event Malfunction Summary report: N

AU2703-03E CLINICAL CHEMISTRY ANALYZER.

MDR report key: 2221141 · Received August 23, 2011

Report

Report Number
2050012-2011-04810
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 28, 2011
Report Date
July 28, 2011
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
JJE
PMA / PMN Number
K003721
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SAMPLES WERE FROM FRESH ADULT PATIENT SAMPLES AND ANALYZED FROM THE PRIMARY TUBES WHICH WERE VACUETTE BD 4 ML TUBES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE SAMPLE PROBE AND THE ISE SAMPLE D-POT ON THE ISE UNIT AS THE SAMPLE ARM, WHICH WAS PREVIOUSLY REPLACED DUE TO MISALIGNMENT, WAS CAUSING GROVES IN THE SAMPLE D-POT. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT AFTER SERVICE WAS PERFORMED FOR A PREVIOUS REPORTABLE EVENT (MDR 2050012-2011-04801) THE (B)(4) CLINICAL CHEMISTRY ANALYZER GENERATED ERRONEOUSLY LOW SODIUM (NA), POTASSIUM (K), AND CHLORIDE (CL) RESULTS ON TWO (2) PATIENT SAMPLES. THE TWO (2) SAMPLES WERE THEN REPEATED AUTOMATICALLY ON THE SAME INSTRUMENT AND GAVE HIGHER RECOVERIES. THE SAMPLES WERE THEN TRANSFERRED TO ANOTHER INSTRUMENT IN THE LABORATORY AND CONFIRMED THE REPEAT RUN VALUES AND THE RESULTS WERE THEN REPORTED OUT OF THE LABORATORY. NO TREATMENT TO PATIENT WAS GIVEN UNTIL RESULTS WERE RELEASED AFTER RETESTING. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AU2703-03E CLINICAL CHEMISTRY ANALYZER. CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER MISHIMA K.K. AU2701-03ISE NA

Patients

Seq Age Sex Outcome Treatment
1