AU2703-03E CLINICAL CHEMISTRY ANALYZER.
Report
- Report Number
- 2050012-2011-04801
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- July 26, 2011
- Report Date
- July 26, 2011
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- JJE
- PMA / PMN Number
- K003721
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE SAMPLE IS SERUM. PER THE CUSTOMER COMPLAINT RECORD, THE INSTRUMENT SAMPLE ARM HAS BEEN REPLACED DUE TO MISALIGNMENT. INVESTIGATION IS STILL PENDING. ADDITIONAL INFORMATION FOR THIS INVESTIGATION IS BEING REPORTED IN MDR 2050012-2011-04810.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT A PAEDIATRIC GLUCOSE SERUM SAMPLE, FROM AN INFANT PATIENT, GENERATED AN ERRONEOUSLY HIGH RESULT OF 4.46 MMOL/L ON THE (B)(4) CLINICAL CHEMISTRY ANALYZER. THE ERRONEOUS RESULT WAS IMMEDIATELY QUERIED BY THE CONSULTANT AS THIS INFANT PATIENT WAS HYPOGLYCEMIC, AND ADDITIONAL ON-GOING MONITORING OF THE INFANT HAD SHOWN THE INFANT TO HAVE LOW GLUCOSE. NO TREATMENT WAS ADMINISTERED BASED ON THE ORIGINAL RESULT. ON REQUEST BY THE HOSPITAL CONSULTANT, THE SAME SERUM SAMPLE WAS RE-TESTED AND CONSISTENT RESULTS OF 1.71MMOL/L AND 1.79 MMOL/L WERE OBTAINED. THE SAME PRIMARY TUBE WAS USED FOR ALL TESTS. PATIENT TREATMENT WAS NOT AFFECTED IN THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AU2703-03E CLINICAL CHEMISTRY ANALYZER. | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER MISHIMA K.K. | AU2701-03ISE | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |