FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2221100 · Received August 23, 2011

Report

Report Number
2531779-2011-06155
Event Type
Malfunction
Date Received
August 23, 2011
Report Date
July 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS. ALL KEYPAD BUTTONS WERE INTERMITTENTLY ACTIVATING PUMP FUNCTIONS WHEN PRESSED. ADHESIVE WAS FOUND UNDER KEYPAD BUTTON CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT FOR THE PAST TWO MONTHS, THE UP, DOWN AND OK BUTTONS FELT LOOSE AND "MUSHY." THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 39 YR