FDA Adverse Event Malfunction Summary report: N

DREAMTOME¿ RX 44

MDR report key: 2221087 · Received August 23, 2011

Report

Report Number
3005099803-2011-02896
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
August 2, 2011
Report Date
August 2, 2011
Manufacturer
BOSTON SCIENTIFIC - SPENCER
Product Code
KNS
PMA / PMN Number
K013153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - THE REPORTED EVENT OF CATHETER SPLIT. THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE CLOSED EXTRUSION AT THE DISTAL TIP WAS RIPPED. THE RIP EXISTED FROM THE POINT WHERE THE GUIDE WIRE CHANNEL ENDS, TO THE TIP, TEARING OPEN THE CLOSED EXTRUSION THROUGH THE DISTAL TIP AND SPLITTING THE TIP. DURING MANUFACTURING, TOME DEVICES ARE 100% INSPECTED SO THE DAMAGE IS LIKELY DUE TO PROCEDURAL FACTORS. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. THE DEVICE HISTORY RECORD REVIEW FOUND THE DEVICE MET ALL MANUFACTURING SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DREAMTOME RX SPHINCTEROTOME WAS USED DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE DISTAL END OF THE CATHETER WAS SPLIT ALL THE WAY TO THE TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER DREAMTOME RX SPHINCTEROTOME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DERMATOME RX SPHINCTEROTOME WAS USED DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE DISTAL END OF THE CATHETER WAS SPLIT ALL THE WAY TO THE TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER DERMATOME RX SPHINCTEROTOME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DREAMTOME¿ RX 44 UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS BOSTON SCIENTIFIC - SPENCER M00584060 14413705

Patients

Seq Age Sex Outcome Treatment
1 65 YR