DREAMTOME¿ RX 44
Report
- Report Number
- 3005099803-2011-02896
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- August 2, 2011
- Report Date
- August 2, 2011
- Manufacturer
- BOSTON SCIENTIFIC - SPENCER
- Product Code
- KNS
- PMA / PMN Number
- K013153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4) - THE REPORTED EVENT OF CATHETER SPLIT. THE DEVICE HAS NOT BEEN RECEIVED FOR ANALYSIS. UPON RECEIPT AND COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND THAT THE CLOSED EXTRUSION AT THE DISTAL TIP WAS RIPPED. THE RIP EXISTED FROM THE POINT WHERE THE GUIDE WIRE CHANNEL ENDS, TO THE TIP, TEARING OPEN THE CLOSED EXTRUSION THROUGH THE DISTAL TIP AND SPLITTING THE TIP. DURING MANUFACTURING, TOME DEVICES ARE 100% INSPECTED SO THE DAMAGE IS LIKELY DUE TO PROCEDURAL FACTORS. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT. THE DEVICE HISTORY RECORD REVIEW FOUND THE DEVICE MET ALL MANUFACTURING SPECIFICATIONS.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DREAMTOME RX SPHINCTEROTOME WAS USED DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE DISTAL END OF THE CATHETER WAS SPLIT ALL THE WAY TO THE TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER DREAMTOME RX SPHINCTEROTOME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DERMATOME RX SPHINCTEROTOME WAS USED DURING A ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THE DISTAL END OF THE CATHETER WAS SPLIT ALL THE WAY TO THE TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER DERMATOME RX SPHINCTEROTOME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DREAMTOME¿ RX 44 | UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) | KNS | BOSTON SCIENTIFIC - SPENCER | M00584060 | 14413705 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |