FDA Adverse Event Malfunction Summary report: N

SMARTDRIVE

MDR report key: 22210228 · Received June 16, 2025

Report

Report Number
3008370857-2025-00143
Event Type
Malfunction
Date Received
June 16, 2025
Date of Event
May 21, 2025
Report Date
August 19, 2025
Manufacturer
MAX MOBILITY LLC
Product Code
ITI
UDI-DI
00861896000310
PMA / PMN Number
K151199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DUE TO AN INCREASE IN REPORTS OF SIMILAR FAILURES, A CAPA INVESTIGATION, 24-015, WAS IMPLEMENTED. INVESTIGATION FOUND AN UNDESIRABLE INTERACTION BETWEEN THE CIRCUIT DESIGN AND THE NEW POTENTIOMETER OVER TIME. FURTHER ENGINEERING ANALYSIS CONCLUDED THAT THE NEW POTENTIOMETER'S INTERNAL CONTACT RESISTANCE WAS CHANGING DURING USE, AND THE CIRCUIT WAS OVERLY SENSITIVE TO THIS CHANGE. THE INVESTIGATION ALSO SHOWED THAT THE INTERNAL CONTACT RESISTANCE OF THE PREVIOUS POTENTIOMETERS, USED PRIOR TO AUGUST 17, 2023, DID NOT CHANGE MUCH OVER TIME WHEN COMPARED TO THE NEW POTENTIOMETER, CONCLUDING THE ROOT-CAUSE FOR THE FAILURE WAS WITHIN THE NEW POTENTIOMETER. AS PART OF THE CAPA, PERMOBIL HAS IMPLEMENTED A MARKET CORRECTION, Z-1116-2025, TO ADDRESS POTENTIALLY AFFECTED COMPONENTS MANUFACTURED BETWEEN AUGUST 17, 2023 THROUGH NOVEMBER 21, 2024. THIS REPORT IS CLAIMED AS BEING THE NEW DESIGN OF SPEED CONTROL DIAL THAT THE END-USER RECEIVED AS PART OF THE MARKET CORRECTION. AT TIME OF THIS SUBMISSION, PERMOBIL HAS YET TO RECEIVE THE SUSPECT COMPONENT FOR INSPECTION, THUS CANNOT CONFIRM THE PRODUCT ALLEGED AS HAVING FAILED TO BE THE NEW DESIGN, NOR CONFIRM A FAILURE HAVING OCCURRED AS ALLEGED. IF PERMOBIL RECEIVES ANY NEW INFORMATION, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE DHR WAS REVIEWED, AND THE DEVICE WAS FOUND TO HAVE MET SPECIFICATION PRIOR TO DISTRIBUTION.

Additional Manufacturer Narrative · 0

SUSPECT COMPONENT WAS RETURNED TO PERMOBIL FOR INSPECTION AND EVALUATION. THROUGH TESTING, PERMOBIL WAS UNABLE TO REPRODUCE THE REPORTED CLAIM OF FAILURE TO STOP. COMPONENT WAS FOUND TO REMAIN FULLY OPERATIONAL WITH NO NOTABLE DEFECTS OR MALFUNCTIONS OCCURRING. PERMOBIL IS UNABLE TO REACH A DETERMINATION AS TO POSSIBLE ROOT CAUSE FOR THE ALLEGED MALFUNCTION WITHOUT SPECULATION.

Description of Event or Problem · 0

RECEIVED REPORT CLAIMING WHILE THE END-USER WAS USING THE SMARTDRIVE MX2+ DEVICE VIA A SPEED CONTROL DIAL, CLAIMS WHILE DIAL IS IN A STAND-BY/NEUTRAL POSITION, ON OCCASION THE DIAL WILL ENGAGE A DRIVE COMMAND WITHOUT PHYSICAL MANIPULATION, AND/OR INCREASE SPEED WITHOUT PHYSICAL MANIPULATION. NO INJURIES WERE REPORTED TO HAVE OCCURRED AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
988250 SMARTDRIVE SMARTDRIVE ITI MAX MOBILITY LLC MX2+ 00861896000310

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown