FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 2221007 · Received August 23, 2011

Report

Report Number
9611451-2011-00510
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 24, 2011
Report Date
July 25, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. (B)(4). THE COMPLAINT RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR VISUAL INSPECTION AND PERFORMANCE TESTING. A HOLE WAS FOUND IN THE RETURNED EVAQUA LIMB, APPROXIMATELY 24CM AWAY FROM THE CONNECTOR ON THE PATIENT END. THE HOLE HAD THE APPEARANCE OF HAVING BEEN PUNCTURED OR SCRATCHED WITH A BLUNT OBJECT. THE PRESSURE TEST REVEALED EXCESSIVE LEAK, DUE TO THE OBSERVED DAMAGE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT NUMBER 110421. BASED ON INSPECTION OF THE HOLE, THE COMPLAINT EVAQUA EXPIRATORY LIMB OF THE BREATHING CIRCUIT WAS PUNCTURED OR SCRATCHED BY A BLUNT OBJECT. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION. ANY BREATHING CIRCUIT WHICH DOES NOT PASS THE PRESSURE TEST IS DISCARDED. THIS SUGGESTS THAT THE BREATHING CIRCUIT WAS PUNCTURED POST-PRODUCTION, POSSIBLY DURING STORAGE OR SETUP. THE KEY DIFFERENCE BETWEEN FPH'S EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY TUBE OF EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED (B)(4), HOWEVER THE EVAQUA TUBING REMAINS MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND CIRCUIT HANGERS. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS; FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE EXPIRATORY TUBE OF AN RT340 BREATHING CIRCUIT WAS LEAKING. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 110421

Patients

Seq Age Sex Outcome Treatment
1