FDA Adverse Event Malfunction Summary report: N

PHOENIX DISPOSABLE PRESSURE TRANSDUCER

MDR report key: 2220902 · Received August 23, 2011

Report

Report Number
2015691-2011-16087
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 24, 2011
Report Date
July 25, 2011
Manufacturer
EDWARDS LIFESCIENCES DR
Product Code
DXO
PMA / PMN Number
K925638
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONE SINGLE DPT - VAMP ADULT KIT WAS RETURNED FOR ANALYSIS. DRY BLOOD WAS VISIBLE AT THE DISTAL TUBING CONNECTOR. THE TUBING WAS COMPLETELY DETACHED FROM THE SAMPLE SITE (Z-SITE) SOLVENT BOND JOINT. INDICATION OF BONDING SOLVENT WAS PRESENT AT A SMALL AREA OF THE TUBING BONDING SURFACE. THE OUTER DIAMETER OF THE TUBING WAS MEASURED NEAR THE POINT OF DETACHMENT AND FOUND WITHIN THE ALLOWABLE SPECIFICATION. ADDITIONALLY, THE SAMPLE SITE INNER DIAMETER WAS MEASURED AND FOUND WITHIN THE ALLOWABLE SPECIFICATION. IT APPEARED THAT THE TUBING DETACHED DUE TO INSUFFICIENT BONDING SOLVENT. VISUAL EXAMINATION WAS PERFORMED UNDER 10X MAGNIFICATION. THE REPORTED DEFECT WAS CONFIRMED TO BE A RELATED TO THE MANUFACTURING PROCESS. THIS ISSUE HAS PREVIOUSLY BEEN INVESTIGATED AND THE ROOT CAUSE WAS DETERMINED TO BE PERSONNEL NOT FOLLOWING WORK INSTRUCTIONS. THE PROCEDURE WAS ALSO REVIEWED TO DETERMINE IF THE WORK INSTRUCTION IS SUFFICIENT; NO DEFICIENCIES WERE IDENTIFIED. AWARENESS TRAINING, INCLUDING REVIEW OF THE CORRECT BONDING PROCESS, WAS PROVIDED TO THE APPLICABLE OPERATORS. BECAUSE THIS LOT NUMBER WAS NOT PROVIDED, IT IS NOT POSSIBLE TO DETERMINE IF THIS UNIT WAS MANUFACTURED BEFORE OR AFTER THE RE-TRAINING.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CONNECTION AT THE DROP PORT (Z-SITE) DISCONNECTED DURING USE. THE REPORTER NOTED THAT THE TUBING PROBABLY DETACHED, RATHER THAN BROKE, BUT THEY WERE NOT CERTAIN. THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHOENIX DISPOSABLE PRESSURE TRANSDUCER TRUWAVE DISPOSABLE PRESSURE TRANSDUCER DXO EDWARDS LIFESCIENCES DR T001671A UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1