FDA Adverse Event Malfunction Summary report: N

FREESTYLE LITE

MDR report key: 2220872 · Received August 23, 2011

Report

Report Number
2954323-2011-04282
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
August 19, 2011
Report Date
November 2, 2011
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW UP REPORT WILL BE FILED ONCE ADDITIONAL INFORMATION IS OBTAINED.

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL FUNCTIONAL TEST RESULTS WERE WITHIN RANGE SPECIFICATION. NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE REPORTED READINGS FOUND IN THE METER'S MEMORY WERE 48 MG/DL AND 211 MG/DL NOT WITH A 10 MINUTE TIMEFRAME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE MONITOR. CUSTOMER REPORTED RECEIVING READINGS OF 48MG/DL, 21 MG/DL AND 211 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN THE VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.