FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2220866 · Received August 23, 2011

Report

Report Number
6000001-2011-20426
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 31, 2011
Report Date
August 8, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CUSTOMER'S REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH F-27 ALARM WAS CONFIRMED AND DUPLICATED DURING PRODUCT EVALUATION. SERVICE DETERMINED THE CAUSE AS A DEFECTIVE SAFETY CLAMP, BUT THE PART IS OBSOLETE AND THE PUMP WAS RETURNED UNREPAIRED. IT WAS RECOMMENDED THAT THE DEVICE NOT BE PUT BACK INTO CLINICAL USE WITHOUT BEING REPAIRED AND TESTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH FAILURE CODE 27. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1