FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III COLONOVIDEOSCOPE

MDR report key: 22208626 · Received June 15, 2025

Report

Report Number
9610595-2025-11229
Event Type
Malfunction
Date Received
June 15, 2025
Date of Event
June 5, 2025
Report Date
June 15, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDF
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND A REPORTABLE MALFUNCTION WAS FOUND. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE J-TUBE FOREIGN OBJECTS WAS TRACED TO COMPONENT FAILURE INDICATING EXPECTED OR RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON J-TUBE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
816486 EVIS EXERA III COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF AIZU OLYMPUS CO., LTD. CF-HQ190I

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown