FDA Adverse Event
Malfunction
Summary report: N
EVIS EXERA III COLONOVIDEOSCOPE
MDR report key: 22208626
·
Received June 15, 2025
Report
- Report Number
- 9610595-2025-11229
- Event Type
- Malfunction
- Date Received
- June 15, 2025
- Date of Event
- June 5, 2025
- Report Date
- June 15, 2025
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDF
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION AND A REPORTABLE MALFUNCTION WAS FOUND. BASED ON THE RESULTS OF THE INVESTIGATION, THE MOST PROBABLE CAUSE OF THE J-TUBE FOREIGN OBJECTS WAS TRACED TO COMPONENT FAILURE INDICATING EXPECTED OR RANDOM COMPONENT FAILURE WITHOUT ANY DESIGN OR MANUFACTURING ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
Description of Event or Problem · 0
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE COLONOVIDEOSCOPE EXHIBITED FOREIGN MATERIAL ON J-TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 816486 | EVIS EXERA III COLONOVIDEOSCOPE | COLONOVIDEOSCOPE | FDF | AIZU OLYMPUS CO., LTD. | CF-HQ190I |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |