COLLEAGUE
Report
- Report Number
- 6000001-2011-20420
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- August 4, 2011
- Report Date
- August 11, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K041191
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).EVALUATION SUMMARY:THE REPORTED CONDITION OF A COLLEAGUE PUMP WITH A FAILURE CODE 570:XXX WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO A USE/USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO FIX THE REPORTED CONDITION.
(B)(4). THE DEVICE WAS EVALUATED ON-SITE AT THE CUSTOMER FACILITY. BAXTER?S INVESTIGATION IS STILL IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH FAIL CODE 570; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.04.00.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |