FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® COMPACT TEST DRUM

MDR report key: 2220680 · Received August 23, 2011

Report

Report Number
1823260-2011-04504
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 22, 2011
Report Date
September 29, 2011
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
K031755
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. DEVICE RECEIVED IN A CONDITION WHICH MADE ANALYSIS IMPOSSIBLE. UNABLE TO TEST BECAUSE AN EMPTY VIAL WAS RETURNED.

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

CALLER REPORTED COMPACT PLUS SYSTEM BLOOD GLUCOSE RESULTS OF 451 MG/DL AND 96 MG/DL WITHIN 10 MINUTES. NO ADVERSE EVENT WAS REPORTED. A REQUEST WAS MADE FOR THE RETURN OF THE METER AND STRIPS, REPLACEMENT SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCU-CHEK ® COMPACT TEST DRUM BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 20729345

Patients

Seq Age Sex Outcome Treatment
1 051 YR ENALAPRIL MALEATE| PAROXETINE| GLYBURIDE