UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-04736
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- March 10, 2011
- Report Date
- March 10, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE SYSTEM WAS ALSO GIVING "CC SAMPLE CUVETTE NO FLUID DETECTED" ERRORS. ON (B)(4) 2011, BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT. THE CUSTOMER STATED TO THE FSE THAT THEY NOTICED WATER BY REAGENT SYRINGE AND REAGENT PROBE B. THE FSE REPLACED THE T-VALVE. THE FSE PRIMED THE INSTRUMENT AND OBSERVED NO LEAKING. AS OF (B)(4) 2011, THE CUSTOMER HAS NOT CONTACTED BECKMAN WITH REQUESTS FOR FURTHER ASSISTANCE FOR THIS ISSUE. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BEC) CONCERNING A FLUID LEAK FROM CARTRIDGE CHEMISTRY (CC) SAMPLE PROBE AND SAMPLE SYRINGE ON UNICEL DXC 800 PRO SYNCHRON CLINICAL SYSTEM. MSDS WAS NOT REVIEWED, BUT THE FACILITY HAS AN EXPOSURE CONTROL PLAN. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION AND THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 PRO SYNCHRON® CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |