FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2220505
·
Received July 18, 2011
Report
- Report Number
- 3008642652-2011-00232
- Event Type
- Malfunction
- Date Received
- July 18, 2011
- Date of Event
- June 23, 2011
- Report Date
- July 15, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGER/MODEM CORD DAMAGED) HAS BEEN CONFIRMED. UPON EVAL, THE POWER BRICK CONNECTOR WAS BENT AND WOULD NOT STAY CONNECTED TO THE CHARGER. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED, BUT WAS LIKELY A RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE POWER BRICK CONNECTOR. THE PT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
THE DAUGHTER OF A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE BATTERY CHARGER CORD WAS DAMAGED. THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |