FDA Adverse Event Other Summary report: N

MAQUET/DATASCOPE

MDR report key: 2220432 · Received August 18, 2011

Report

Report Number
MW5021799
Event Type
Other
Date Received
August 18, 2011
Date of Event
August 8, 2011
Report Date
August 10, 2011
Manufacturer
DATASCOPE CORP
Product Code
DSP
Report Source
Voluntary report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

ON (B)(6) 2011 (B)(4): PT ADMITTED POSSIBLE MI, ACUTE PULMONARY EDEMA AND INTUBATION. UNDERWENT AN EMERGENT CARDIAC CATHETERIZATION, IABP, MECHANICAL VENTILATION AND TRANSFERRED TO ICU POST PROCEDURE. ON (B)(6) 2011 (B)(4): THE IABP AUDIBLE ALARM SOUNDS OFF AND "LOW HELIUM" MESSAGE IS NOTED. NURSE REVIEWS THE "HELP" SCREEN ALARM/ADVISORY. THE CORRECTIVE ACTIONS ON THE HELP SCREEN RECOMMENDS TO OPEN HELIUM TANK AND REPLACE THE HELIUM TANK. RN PROCEEDED TO CONTACT THE PERFUSIONIST TO HELP REPLACE THE TANK AND FAILED TO CHECK THAT THE HELIUM TANK WAS OPEN (HELIUM TANK WAS CLOSED). CONCURRENTLY, PT WENT ASYSTOLE AND CODE BLUE WAS CALLED. RESUSCITATION EFFORTS WERE UNSUCCESSFUL AND PT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MAQUET/DATASCOPE DATASCOPE IABP DSP DATASCOPE CORP CS300

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other