FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2220414
·
Received August 23, 2011
Report
- Report Number
- 2531779-2011-06114
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- July 22, 2011
- Report Date
- July 25, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE OK BUTTON WAS STICKING AND CONTINUING TO SCROLL. THE PATIENT REPORTED THAT THERE WAS NO DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP IN AN UNDERGARMENT AND CLEANED THE PUMP WITH ALCOHOL; CUSTOMER SUPPORT ADVISED THE PATIENT NOT TO USE SOLVENTS ON THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 1250 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 21 YR |