ECHELON*FLEX60 ARTICULATING

MDR report key: 2220410 · Received August 23, 2011

Report

Report Number
3005075853-2011-03462
Event Type
Malfunction
Date Received
August 23, 2011
Report Date
July 29, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO CARTRIDGE RELOAD LOADED ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP RECTUM PROCEDURE, AFTER RELOADING THERE WAS AN INCOMPLETE STAPLE LINE. NO FURTHER INFORMATION IS AVAILABLE. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON*FLEX60 ARTICULATING STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK

Patients

Seq Age Sex Outcome Treatment