FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2220362 · Received August 23, 2011

Report

Report Number
2134265-2011-03451
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 29, 2011
Report Date
July 29, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE EVALUATION BY MANUFACTURER: THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREP FOR A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, A BALLOON RUPTURE OCCURRED. WHILE AIR BLEEDING, THE 3.25MM X 15MM NC QUANTUM APEX MR BALLOON RUPTURED. NO PATIENT COMPLICATIONS WERE REPORTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE PATIENT STATUS WAS RECORDED AS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415320 14048507

Patients

Seq Age Sex Outcome Treatment
1